The ac-9 Plus combines the power of the standard ac-9 dual
path absorption and attenuation meter with the added
capability of accepting and integrating data from the WET
Labs’ ECO-VSF sensors as well as a CTD. The data from
these ancillary sensors is incorporated into the serial (RS232) data stream. WETView, our host data collection and
real-time display program, records and displays in real-time
the data from the ac-9 Plus, as well as conductivity, water
temperature, depth and optical backscattering. This powerful
new tool will allow researchers to simultaneously collect the
data needed to describe inherent optical property
measurements of natural waters for effects such as salinity
and temperature with one integrated package. It will also
provide important information about the Volume Scattering
Function (VSF). The ac-9 Plus is available in several
standard configurations: the ac-9 Plus + CTD, ac-9 Plus +
ECO-VSF + CTD, and ac-9 Plus + CTD + ECO-VSF 3. It
supports many popular CTDs including the SBE 19, 25, and
37, as well as CTDs from other manufacturers. Upgrades are
available for existing ac-9 sensors.
Specification |
Diameter | 10.4 cm |
Spectral range |
410 ~ 715 nm |
Length |
76.5 cm |
Band pass |
10 nm / channel |
Weight |
6.6 kg in air
2.0 kg in water |
Pathlength |
10 or 25 cm |
Pressure housing |
Acetal copolymer |
Beam cross-section |
8 nm dia. |
Ele. input | 10 ~ 16 VDC |
Linearity |
> 99 % R2 |
Current draw | 50mA |
Output wavelengths |
9 |
Serial output | RS-232, 485 |
Accuracy |
0.001m-1 |
Connector | MCBH6M |
Dynamic range |
0.001 ~ 10 m-1 |
Sample rate | 6 scans/sec |
Temperature range |
0 ~ 30 degree C |
Pressure sensor | optional |
Depth rating |
500 m |
|
|